VECTOR™ PYROELECTRIC DETECTOR SPECIFICATIONS:
Model
P 25 PHF 25 PHD 25 PHDX 25 PHDX25UV SP 25 SPHF 25 SPHD 25
Aperture Diameter 25.4 mm 25.4 mm 25.4 mm 7mm 7mm 25.4 mm 25.4 mm 25.4 mm
Spectral Response .193-10.6µm.4-2µm .193-2µm .193-10.6µm
Maximum Average Power
Minimum Energy
Noise Equivalent Energy 4 µJ
Maximum Energy Density Note 2 Note 3 Note 4 Note 2
Peak Power Density
100MW/cm
2
800MW/cm
2
400MW/cm
2
100MW/cm
2
Accuracy 5% 5% 8%^ 8%^ 8%^ 5% 5% 8%^
Output Sensitivity 8 V/J 8 V/J 2 V/J 2 V/J 2 V/J 8 V/J 8 V/J 2 V/J
Maximum Repetition Rate 100 pps 400 pps 40 pps 40 pps 40 pps 100 pps 400 pps 40 pps
Maximum Pulse Duration 0.2 msec 0.045 msec 0.2 msec 0.2 msec 0.2 msec 0.2 msec 0.045 msec 0.2 msec
Dimensions D x L - inches 2.4 x 2.3 2.4 x 2.3 2.4 x 2.3 2.4 x 3.9 2.4 x 3.9 2.3x2.3x0.6 2.3x2.3x0.6 2.3x2.3x0.6
cm 6.1 x 5.8 6.1 x 5.8 6.1 x 5.8 6.1 x 9.9 6.1 x 9.9 5.8x5.8x1.4 5.8x5.8x1.4 5.8x5.8x1.4
Weight - pounds/kgs 0.9/1.4 0.9/1.4 0.9/1.4 1.1/0.5 1.1/0.5 0.3/0.14 0.3/0.14 0.3/0.14
Indicator Compatibility H410, H410D, S310, S310D, D200PC, D200P
Model
P 50 PHF 50 PHD 50 PHDX 50 PHDX50UV SP 50 SPHF 50 SPHD 50
Maximum Beam Diameter 50.8 mm 50.8 mm 50.8 mm 15 mm 15 mm 50.8 mm 50.8 mm 50.8 mm
Spectral Response .193-10.6µm.4-2µm .193-2µm .193-10.6µm
Maximum Average Power
Minimum Energy
Noise Equivalent Energy
Maximum Energy Density Note 2 Note 3 Note 4 Note 2
Peak Power Density
100MW/cm
2
800MW/cm
2
400MW/cm
2
100MW/cm
2
Accuracy 5% 5% 8%^ 8%^ 8%^ 5% 5% 8%^
Output Sensitivity
Maximum Repetition Rate 50 pps 400 pps 20 pps 20 pps 20 pps 50 pps 400 pps 20 pps
Maximum Pulse Duration 0.4 msec 0.045 msec 0.4 msec 0.4 msec 0.4 msec 0.4 msec 0.045 msec 0.4 msec
Dimensions D x L - inches 3.5 x 2.3 3.5 x 2.3 3.5 x 2.3 3.5 x 3.9 3.5 x 3.9 3x3x0.6 3x3x0.6 3x3x0.6
cm 8.8 x 5.8 8.8 x 5.8 8.8 x 5.8 8.8 x 9.9 8.8 x 9.9 7.6x7.6x1.5 7.6x7.6x1.5 7.6x7.6x1.5
Weight - pounds/kgs 1.5/0.68 1.5/0.68 1.5/0.68 1.7/0.77 1.7/0.77 0.4/0.18 0.4/0.18 0.4/0.18
Indicator Compatibility H410, H410D, S310, S310D, D200PC, D200P
^Beam centered on absorber
Note 1: Max J/cm2 =316 x (pulse width)1/2
Note 2: HD models Max J/cm
2
= 4500 x (pulse width)
1/2
to a maximum of 1.4 J/cm
2
.
Maximum pulse width of the pyroelectric detector must be observed.
Note 3: HDX models Max J/cm
2
= 36,000 x (pulse width)
1/2
to a maximum of 12.6 J/cm
2
.
Maximum pulse width of the pyroelectric detector must be observed.
Note 4: HDXUV models Max J/cm
2
= 18,000 x (pulse width)
1/2
to a maximum of 5.6 J/cm
2
.
Maximum pulse width of the pyroelectric detector must be observed.
5 W with full illumination of the detector
10 W with full illumination of the detector
.193-26µm .193-26µm
.193-26µm .193-26µm
7% of selected range
n/a n/a
n/a n/a
2 V/J
Note 1 Note 1
Note 1 Note 1
16 µJ
7% of selected range
CE MARK CERTIFICATION:
All of the pyroelectric detectors listed in this manual have been certified for the European CE mark.
ENVIRONMENTAL REQUIREMENTS:
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance with
IEC 664 and transient overvoltages according to Installation Categories (Overvoltage Categories) II. Note that
each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated Electromagnetic Field
Requirements. The system, meter and detector, is designed to measure radiation within the test's radiation band.
The detectors were held outside the radiated electromagnetic field during this test. It is up to the user to be
aware of RF fields present during measurements and their effects if any on those measurements.