Pyroelectric Detector
Setup and Operating
Procedures
PN10354M
Vista de página 0
1 2 3 4 5 6 ... 15 16

Resumo do Conteúdo

Página 1 -

Pyroelectric Detector Setup and Operating Procedures PN10354M

Página 2 - Table of Contents:

10 Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.855 89.26 3.233 88.79 3.725 88.11 4.394 87

Página 3 - ENVIRONMENTAL REQUIREMENTS:

11Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%5.357 86.95 6.682 86.05 8.879 84.32 13.09 80.

Página 4 - UNPACKING AND SET UP:

12HF MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile high frequency (HF) model pyroelectric detec

Página 5 - DIFFERENT WAVELENGTHS:

13Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.067 60.27 2.257 60.06 2.487 59.79 2.768 59.

Página 6 - LIMITED WARRANTY:

14Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%3.120 59.47 3.576 59.37 4.188 58.92 5.052 58.

Página 7

15Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%6.365 57.97 7.870 57.74 10.31 60.33 14.93 43.

Página 8

16P MODEL PYROELECTRIC DETECTOR ABSORPTION VS. WAVELENGTH: Use this table for standard and slim profile painted (P) model pyroelectric detectors. Wav

Página 9

2Thank you for choosing a Scientech Vector pyroelectric detector. Scientech, an ISO 9001 registered company, and our employees are pleased to provid

Página 10

3VECTOR™ PYROELECTRIC DETECTOR SPECIFICATIONS: ModelP 25 PHF 25 PHD 25 PHDX 25 PHDX25UV SP 25 SPHF 25 SPHD 25Aperture Diameter 25.4 mm 25.4 mm 25.4 m

Página 11

4ABSORPTION OF HD ABSORBING MATERIAL: Warning: You must exercise caution when using HD detectors. They exhibit spectral reflection of between 7% an

Página 12

5RELATIVE SPECTRAL RESPONSE20304050607080901000.1 1 10 100WAVELENGTH [microns]PERCENT ABSORPTIONP and SPHR-PPHD, PHDX* and SPHDHR-PHFPHF and SPHF*HDX

Página 13

6FACTORY RECALIBRATION: Scientech recommends that a complete calibration be performed annually to verify system accuracy. Please contact our Product

Página 14

7HD PYROELECTRIC ABSORPTION VS. WAVELENGTH: Use this table for all standard and slim profile HD model pyroelectric detectors. Wavelength Absorption W

Página 15

8Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%1.245 91.78 1.510 91.70 1.775 90.84 2.016 90.8

Página 16

9Wavelength Absorption Wavelength Absorption Wavelength Absorption Wavelength Absorptionµm% µm% µm% µm%2.129 90.76 2.255 90.17 2.390 89.92 2.557 89.3